Chapter IV: Part II: Volumetric Analysis (1)
The processes of volumetric analysis are, in general, simpler than those of gravimetric analysis and accordingly serve best as an introduction to the practice of quantitative analysis. For their execution there are required, first, an accurate balance with which to weigh the material for analysis; second, graduated instruments in which to measure the volume of the solutions employed; third, standard solutions, that is, solutions the value of which is accurately known; and fourth, indicators, which will furnish accurate evidence of the point at which the desired reaction is completed. The nature of the indicators employed will be explained in connection with the different analyses.
The process whereby a !standard solution! is brought into reaction is called !titration!, and the point at which the reaction is exactly completed is called the !end-point!. The !indicator! should show the !end-point! of the !titration!. The volume of the standard solution used then furnishes the measure of the substance to be determined as truly as if that substance had been separated and weighed.
The processes of volumetric analysis are easily classified, according to their character, into:
I. NEUTRALIZATION METHODS; such, for example, as those of acidimetry and alkalimetry.
II. OXIDATION PROCESSES; as exemplified in the determination of ferrous iron by its oxidation with potassium bichromate.
III. PRECIPITATION METHODS; of which the titration for silver with potassium thiocyanate solution is an illustration.
From a somewhat different standpoint the methods in each case may be subdivided into (a) DIRECT METHODS, in which the substance to be measured is directly determined by titration to an end-point with a standard solution; and (b) INDIRECT METHODS, in which the substance itself is not measured, but a quantity of reagent is added which is known to be an excess with respect to a specific reaction, and the unused excess determined by titration. Examples of the latter class will be pointed out as they occur in the procedures.
MEASURING INSTRUMENTS
THE ANALYTICAL BALANCE
For a complete discussion of the physical principles underlying the construction and use of balances, and the various methods of weighing, the student is referred to larger manuals of Quantitative Analysis, such as those of Fresenius, or Treadwell-Hall, and particularly to the admirable discussion of this topic in Morse's !Exercises in Quantitative Chemistry!.
The statements and rules of procedure which follow are sufficient for the intelligent use of an analytical balance in connection with processes prescribed in this introductory manual. It is, however, imperative that the student should make himself familiar with these essential features of the balance, and its use. He should fully realize that the analytical balance is a delicate instrument which will render excellent service under careful treatment, but such treatment is an essential condition if its accuracy is to be depended upon. He should also understand that no set of rules, however complete, can do away with the necessity for a sense of personal responsibility, since by carelessness he can render inaccurate not only his own analyses, but those of all other students using the same balance.
Before making any weighings the student should seat himself before a balance and observe the following details of construction:
1. The balance case is mounted on three brass legs, which should preferably rest in glass cups, backed with rubber to prevent slipping. The front legs are adjustable as to height and are used to level the balance case; the rear leg is of permanent length.
2. The front of the case may be raised to give access to the balance. In some makes doors are provided also at the ends of the balance case.
3. The balance beam is mounted upon an upright in the center of the case on the top of which is an inlaid agate plate. To the center of the beam there is attached a steel or agate knife-edge on which the beam oscillates when it rests on the agate plate.
4. The balance beam, extending to the right and left, is graduated along its upper edge, usually on both sides, and has at its extremities two agate or steel knife-edges from which are suspended stirrups. Each of these stirrups has an agate plate which, when the balance is in action, rests upon the corresponding knife-edge of the beam. The balance pans are suspended from the stirrups.
5. A pointer is attached to the center of the beam, and as the beam oscillates this pointer moves in front of a scale near the base of the post.
6. At the base of the post, usually in the rear, is a spirit-level.
7. Within the upright is a mechanism, controlled by a knob at the front of the balance case, which is so arranged as to raise the entire beam slightly above the level at which the knife-edges are in contact with the agate plates. When the balance is not in use the beam must be supported by this device since, otherwise, the constant jarring to which a balance is inevitably subjected, will soon dull the knife-edges, and lessen the sensitiveness of the balance.
8. A small weight, or bob, is attached to the pointer (or sometimes to the beam) by which the center of gravity of the beam and its attachments may be regulated. The center of gravity must lie very slightly below the level of the agate plates to secure the desired sensitiveness of the balance. This is provided for when the balance is set up and very rarely requires alteration. The student should never attempt to change this adjustment.
9. Below the balance pans are two pan-arrests operated by a button from the front of the case. These arrests exert a very slight upward pressure upon the pans and minimize the displacement of the beam when objects or weights are being placed upon the pans.
10. A movable rod, operated from one end of the balance case, extends over the balance beam and carries a small wire weight, called a rider. By means of this rod the rider can be placed upon any desired division of the scale on the balance beam. Each numbered division on the beam corresponds to one milligram, and the use of the rider obviates the placing of very small fractional weights on the balance pan.
If a new rider is purchased, or an old one replaced, care must be taken that its weight corresponds to the graduations on the beam of the balance on which it is to be used. The weight of the rider in milligrams must be equal to the number of large divisions (5, 6, 10, or 12) between the central knife-edge and the knife-edge at the end of the beam. It should be noted that on some balances the last division bears no number. Each new rider should be tested against a 5 or 10-milligram weight.
In some of the most recent forms of the balance a chain device replaces the smaller weights and the use of the rider as just described.
Before using a balance, it is always best to test its adjustment. This is absolutely necessary if the balance is used by several workers; it is always a wise precaution under any conditions. For this purpose, brush off the balance pans with a soft camel's hair brush. Then note (1) whether the balance is level; (2) that the mechanism for raising and lowering the beams works smoothly; (3) that the pan-arrests touch the pans when the beam is lowered; and (4) that the needle swings equal distances on either side of the zero-point when set in motion without any load on the pans. If the latter condition is not fulfilled, the balance should be adjusted by means of the adjusting screw at the end of the beam unless the variation is not more than one division on the scale; it is often better to make a proper allowance for this small zero error than to disturb the balance by an attempt at correction. Unless a student thoroughly understands the construction of a balance he should never attempt to make adjustments, but should apply to the instructor in charge.
The object to be weighed should be placed on the left-hand balance pan and the weights upon the right-hand pan. Every substance which could attack the metal of the balance pan should be weighed upon a watch-glass, and all objects must be dry and cold. A warm body gives rise to air currents which vitiate the accuracy of the weighing.
The weights should be applied in the order in which they occur in the weight-box (not at haphazard), beginning with the largest weight which is apparently required. After a weight has been placed upon the pan the beam should be lowered upon its knife-edges, and, if necessary, the pan-arrests depressed. The movement of the pointer will then indicate whether the weight applied is too great or too small. When the weight has been ascertained, by the successive addition of small weights, to the nearest 5 or 10 milligrams, the weighing is completed by the use of the rider. The correct weight is that which causes the pointer to swing an equal number of divisions to the right and left of the zero-point, when the pointer traverses not less than five divisions on either side.
The balance case should always be closed during the final weighing, while the rider is being used, to protect the pans from the effect of air currents.
Before the final determination of an exact weight the beam should always be lifted from the knife-edges and again lowered into place, as it frequently happens that the scale pans are, in spite of the pan-arrests, slightly twisted by the impact of the weights, the beam being thereby virtually lengthened or shortened. Lifting the beam restores the proper alignment.
The beam should never be set in motion by lowering it forcibly upon the knife-edges, nor by touching the pans, but rather by lifting the rider (unless the balance be provided with some of the newer devices for the purpose), and the swing should be arrested only when the needle approaches zero on the scale, otherwise the knife-edges become dull. For the same reason the beam should never be left upon its knife-edges, nor should weights be removed from or placed on the pans without supporting the beam, except in the case of the small fractional weights.
When the process of weighing has been completed, the weight should be recorded in the notebook by first noting the vacant spaces in the weight-box, and then checking the weight by again noting the weights as they are removed from the pan. This practice will often detect and avoid errors. It is obvious that the weights should always be returned to their proper places in the box, and be handled only with pincers.
It should be borne in mind that if the mechanism of a balance is deranged or if any substance is spilled upon the pans or in the balance case, the damage should be reported at once. In many instances serious harm can be averted by prompt action when delay might ruin the balance.
Samples for analysis are commonly weighed in small tubes with cork stoppers. Since the stoppers are likely to change in weight from the varying amounts of moisture absorbed from the atmosphere, it is necessary to confirm the recorded weight of a tube which has been unused for some time before weighing out a new portion of substance from it.
WEIGHTS
The sets of weights commonly used in analytical chemistry range from 20 grams to 5 milligrams. The weights from 20 grams to 1 gram are usually of brass, lacquered or gold plated. The fractional weights are of German silver, gold, platinum or aluminium. The rider is of platinum or aluminium wire.
The sets of weights purchased from reputable dealers are usually sufficiently accurate for analytical work. It is not necessary that such a set should be strictly exact in comparison with the absolute standard of weight, provided they are relatively correct among themselves, and provided the same set of weights is used in all weighings made during a given analysis. The analyst should assure himself that the weights in a set previously unfamiliar to him are relatively correct by a few simple tests. For example, he should make sure that in his set two weights of the same denomination (i.e., two 10-gram weights, or the two 100-milligram weights) are actually equal and interchangeable, or that the 500-milligram weight is equal to the sum of the 200, 100, 100, 50, 20, 20 and 10-milligram weights combined, and so on. If discrepancies of more than a few tenths of a milligram (depending upon the total weight involved) are found, the weights should be returned for correction. The rider should also be compared with a 5 or 10-milligram weight.
In an instructional laboratory appreciable errors should be reported to the instructor in charge for his consideration.
When the highest accuracy is desired, the weights may be calibrated and corrections applied. A calibration procedure is described in a paper by T.W. Richards, !J. Am. Chem. Soc.!, 22, 144, and in many large text-books.
Weights are inevitably subject to corrosion if not properly protected at all times, and are liable to damage unless handled with great care. It is obvious that anything which alters the weight of a single piece in an analytical set will introduce an error in every weighing made in which that piece is used. This source of error is often extremely obscure and difficult to detect. The only safeguard against such errors is to be found in scrupulous care in handling and protection on the part of the analyst, and an equal insistence that if several analysts use the same set of weights, each shall realize his responsibility for the work of others as well as his own.
BURETTES
A burette is made from a glass tube which is as uniformly cylindrical as possible, and of such a bore that the divisions which are etched upon its surface shall correspond closely to actual contents.
The tube is contracted at one extremity, and terminates in either a glass stopcock and delivery-tube, or in such a manner that a piece of rubber tubing may be firmly attached, connecting a delivery-tube of glass. The rubber tubing is closed by means of a glass bead. Burettes of the latter type will be referred to as "plain burettes."
The graduations are usually numbered in cubic centimeters, and the latter are subdivided into tenths.
One burette of each type is desirable for the analytical procedures which follow.
PREPARATION OF A BURETTE FOR USE
The inner surface of a burette must be thoroughly cleaned in order that the liquid as drawn out may drain away completely, without leaving drops upon the sides. This is best accomplished by treating the inside of the burette with a warm solution of chromic acid in concentrated sulphuric acid, applied as follows: If the burette is of the "plain" type, first remove the rubber tip and force the lower end of the burette into a medium-sized cork stopper. Nearly fill the burette with the chromic acid solution, close the upper end with a cork stopper and tip the burette backward and forward in such a way as to bring the solution into contact with the entire inner surface. Remove the stopper and pour the solution into a stock bottle to be kept for further use, and rinse out the burette with water several times. Unless the water then runs freely from the burette without leaving drops adhering to the sides, the process must be repeated (Note 1).
If the burette has a glass stopcock, this should be removed after the cleaning and wiped, and also the inside of the ground joint. The surface of the stopcock should then be smeared with a thin coating of vaseline and replaced. It should be attached to the burette by means of a wire, or elastic band, to lessen the danger of breakage.
Fill the burettes with distilled water, and allow the water to run out through the stopcock or rubber tip until convinced that no air bubbles are inclosed (Note 2). Fill the burette to a point above the zero-point and draw off the water until the meniscus is just below that mark. It is then ready for calibration.
[Note 1: The inner surface of the burette must be absolutely clean if the liquid is to run off freely. Chromic acid in sulphuric acid is usually found to be the best cleansing agent, but the mixture must be warm and concentrated. The solution can be prepared by pouring over a few crystals of potassium bichromate a little water and then adding concentrated sulphuric acid.]
[Note 2: It is always necessary to insure the absence of air bubbles in the tips or stopcocks. The treatment described above will usually accomplish this, but, in the case of plain burettes it is sometimes better to allow a little of the liquid to flow out of the tip while it is bent upwards. Any air which may be entrapped then rises with the liquid and escapes.
If air bubbles escape during subsequent calibration or titration, an error is introduced which vitiates the results.]
READING OF A BURETTE
All liquids when placed in a burette form what is called a meniscus at their upper surfaces. In the case of liquids such as water or aqueous solutions this meniscus is concave, and when the liquids are transparent accurate readings are best obtained by observing the position on the graduated scales of the lowest point of the meniscus. This can best be done as follows: Wrap around the burette a piece of colored paper, the straight, smooth edges of which are held evenly together with the colored side next to the burette (Note 1). Hold the paper about two small divisions below the meniscus and raise or lower the level of the eyes until the edge of the paper at the back of the burette is just hidden from the eye by that in front (Note 2). Note the position of the lowest point of the curve of the meniscus, estimating the tenths of the small divisions, thus reading its position to hundredths of a cubic centimeter.
[Note 1: The ends of the colored paper used as an aid to accurate readings may be fastened together by means of a gummed label. The paper may then remain on the burette and be ready for immediate use by sliding it up or down, as required.]
[Note 2: To obtain an accurate reading the eye must be very nearly on a level with the meniscus. This is secured by the use of the paper as described. The student should observe by trial how a reading is affected when the meniscus is viewed from above or below.
The eye soon becomes accustomed to estimating the tenths of the divisions. If the paper is held as directed, two divisions below the meniscus, one whole division is visible to correct the judgment. It is not well to attempt to bring the meniscus exactly to a division mark on the burette. Such readings are usually less accurate than those in which the tenths of a division are estimated.]
CALIBRATION OF GLASS MEASURING DEVICES
If accuracy of results is to be attained, the correctness of all measuring instruments must be tested. None of the apparatus offered for sale can be implicitly relied upon except those more expensive instruments which are accompanied by a certificate from the !National Bureau of Standards! at Washington, or other equally authentic source.
The bore of burettes is subject to accidental variations, and since the graduations are applied by machine without regard to such variations of bore, local errors result.
The process of testing these instruments is called !calibration!. It is usually accomplished by comparing the actual weight of water contained in the instrument with its apparent volume.
There is, unfortunately, no uniform standard of volume which has been adopted for general use in all laboratories. It has been variously proposed to consider the volume of 1000 grams of water at 4°, 15.5°, 16°, 17.5°, and even 20°C., as a liter for practical purposes, and to consider the cubic centimeter to be one one-thousandth of that volume. The true liter is the volume of 1000 grams of water at 4°C.; but this is obviously a lower temperature than that commonly found in laboratories, and involves the constant use of corrections if taken as a laboratory standard. Many laboratories use 15.5°C. (60° F.) as the working standard. It is plain that any temperature which is deemed most convenient might be chosen for a particular laboratory, but it cannot be too strongly emphasized that all measuring instruments, including burettes, pipettes, and flasks, should be calibrated at that temperature in order that the contents of each burette, pipette, etc., shall be comparable with that of every other instrument, thus permitting general interchange and substitution. For example, it is obvious that if it is desired to remove exactly 50 cc. from a solution which has been diluted to 500 cc. in a graduated flask, the 50 cc. flask or pipette used to remove the fractional portion must give a correct reading at the same temperature as the 500 cc. flask. Similarly, a burette used for the titration of the 50 cc. of solution removed should be calibrated under the same conditions as the measuring flasks or pipettes employed with it.
The student should also keep constantly in mind the fact that all volumetric operations, to be exact, should be carried out as nearly at a constant temperature as is practicable. The spot selected for such work should therefore be subject to a minimum of temperature variations, and should have as nearly the average temperature of the laboratory as is possible. In all work, whether of calibration, standardization, or analysis, the temperature of the liquids employed must be taken into account, and if the temperature of these liquids varies more than 3° or 4° from the standard temperature chosen for the laboratory, corrections must be applied for errors due to expansion or contraction, since volumes of a liquid measured at different times are comparable only under like conditions as to temperature. Data to be used for this purpose are given in the Appendix. Neglect of this correction is frequently an avoidable source of error and annoyance in otherwise excellent work. The temperature of all solutions at the time of standardization should be recorded to facilitate the application of temperature corrections, if such are necessary at any later time.
CALIBRATION OF THE BURETTES
Two burettes, one at least of which should have a glass stopper, are required throughout the volumetric work. Both burettes should be calibrated by the student to whom they are assigned.
PROCEDURE.--Weigh a 50 cc., flat-bottomed flask (preferably a light-weight flask), which must be dry on the outside, to the nearest centigram. Record the weight in the notebook. (See Appendix for suggestions as to records.) Place the flask under the burette and draw out into it about 10 cc. of water, removing any drop on the tip by touching it against the inside of the neck of the flask. Do not attempt to stop exactly at the 10 cc. mark, but do not vary more than 0.1 cc. from it. Note the time, and at the expiration of three minutes (or longer) read the burette accurately, and record the reading in the notebook (Note 1). Meanwhile weigh the flask and water to centigrams and record its weight (Note 2). Draw off the liquid from 10 cc. to about 20 cc. into the same flask without emptying it; weigh, and at the expiration of three minutes take the reading, and so on throughout the length of the burette. When it is completed, refill the burette and check the first calibration.
The differences in readings represent the apparent volumes, the differences in weights the true volumes. For example, if an apparent volume of 10.05 cc. is found to weigh 10.03 grams, it may be assumed with sufficient accuracy that the error in that 10 cc. amounts to -0.02 cc., or -0.002 for each cubic centimeter (Note 3).
In the calculation of corrections the temperature of the water must be taken into account, if this varies more than 4°C. from the laboratory standard temperature, consulting the table of densities of water in the Appendix.
From the final data, plot the corrections to be applied so that they may be easily read for each cubic centimeter throughout the burette. The total correction at each 10 cc. may also be written on the burette with a diamond, or etching ink, for permanence of record.
[Note 1: A small quantity of liquid at first adheres to the side of even a clean burette. This slowly unites with the main body of liquid, but requires an appreciable time. Three minutes is a sufficient interval, but not too long, and should be adopted in every instance throughout the whole volumetric practice before final readings are recorded.]
[Note 2: A comparatively rough balance, capable of weighing to centigrams, is sufficiently accurate for use in calibrations, for a moment's reflection will show that it would be useless to weigh the water with an accuracy greater than that of the readings taken on the burette. The latter cannot exceed 0.01 cc. in accuracy, which corresponds to 0.01 gram.
The student should clearly understand that !all other weighings!, except those for calibration, should be made accurately to 0.0001 gram, unless special directions are given to the contrary.
Corrections for temperature variations of less than 4°C. are negligible, as they amount to less than 0.01 gram for each 10 grams of water withdrawn.]
[Note 3: Should the error discovered in any interval of 10 cc. on the burette exceed 0.10 cc., it is advisable to weigh small portions (even 1 cc.) to locate the position of the variation of bore in the tube rather than to distribute the correction uniformly over the corresponding 10 cc. The latter is the usual course for small corrections, and it is convenient to calculate the correction corresponding to each cubic centimeter and to record it in the form of a table or calibration card, or to plot a curve representing the values.
Burettes may also be calibrated by drawing off the liquid in successive portions through a 5 cc. pipette which has been accurately calibrated, as a substitute for weighing. If many burettes are to be tested, this is a more rapid method.]
PIPETTES
A !pipette! may consist of a narrow tube, in the middle of which is blown a bulb of a capacity a little less than that which it is desired to measure by the pipette; or it may be a miniature burette, without the stopcock or rubber tip at the lower extremity. In either case, the flow of liquid is regulated by the pressure of the finger on the top, which governs the admission of the air.
Pipettes are usually already graduated when purchased, but they require calibration for accurate work.
CALIBRATION OF PIPETTES
PROCEDURE.--Clean the pipette. Draw distilled water into it by sucking at the upper end until the water is well above the graduation mark. Quickly place the forefinger over the top of the tube, thus preventing the entrance of air and holding the water in the pipette. Cautiously admit a little air by releasing the pressure of the finger, and allow the level of the water to fall until the lowest point of the meniscus is level with the graduation. Hold the water at that point by pressure of the finger and then allow the water to run out from the pipette into a small tared, or weighed, beaker or flask. After a definite time interval, usually two to three minutes, touch the end of the pipette against the side of the beaker or flask to remove any liquid adhering to it (Note 1). The increase in weight of the flask in grams represents the volume of the water in cubic centimeters delivered by the pipette. Calculate the necessary correction.
[Note 1: A definite interval must be allowed for draining, and a definite practice adopted with respect to the removal of the liquid which collects at the end of the tube, if the pipette is designed to deliver a specific volume when emptied. This liquid may be removed at the end of a definite interval either by touching the side of the vessel or by gently blowing out the last drops. Either practice, when adopted, must be uniformly adhered to.]
FLASKS
!Graduated or measuring flasks! are similar to the ordinary flat-bottomed flasks, but are provided with long, narrow necks in order that slight variations in the position of the meniscus with respect to the graduation shall represent a minimum volume of liquid. The flasks must be of such a capacity that, when filled with the specified volume, the liquid rises well into the neck.
GRADUATION OF FLASKS
It is a general custom to purchase the flasks ungraduated and to graduate them for use under standard conditions selected for the laboratory in question. They may be graduated for "contents" or "delivery." When graduated for "contents" they contain a specified volume when filled to the graduation at a specified temperature, and require to be washed out in order to remove all of the solution from the flask. Flasks graduated for "delivery" will deliver the specified volume of a liquid without rinsing. A flask may, of course, be graduated for both contents and delivery by placing two graduation marks upon it.
PROCEDURE.--To calibrate a flask for !contents!, proceed as follows: Clean the flask, using a chromic acid solution, and dry it carefully outside and inside. Tare it accurately; pour water into the flask until the weight of the latter counterbalances weights on the opposite pan which equal in grams the number of cubic centimeters of water which the flask is to contain. Remove any excess of water with the aid of filter paper (Note 1). Take the flask from the balance, stopper it, place it in a bath at the desired temperature, usually 15.5° or 17.5°C., and after an hour mark on the neck with a diamond the location of the lowest point of the meniscus (Note 2). The mark may be etched upon the flask by hydrofluoric acid, or by the use of an etching ink now commonly sold on the market.
To graduate a flask which is designed to !deliver! a specified volume, proceed as follows: Clean the flask as usual and wipe all moisture from the outside. Fill it with distilled water. Pour out the water and allow the water to drain from the flask for three minutes. Counterbalance the flask with weights to the nearest centigram. Add weights corresponding in grams to the volume desired, and add distilled water to counterbalance these weights. An excess of water, or water adhering to the neck of the flask, may be removed by means of a strip of clean filter paper. Stopper the flask, place it in a bath at 15.5°C. or 17.5°C. and, after an hour, mark the location of the lowest point of the meniscus, as described above.
[Note 1: The allowable error in counterbalancing the water and weights varies with the volume of the flask. It should not exceed one ten-thousandth of the weight of water.]
[Note 2: Other methods are employed which involve the use of calibrated apparatus from which the desired volume of water may be run into the dry flask and the position of the meniscus marked directly upon it. For a description of a procedure which is most convenient when many flasks are to be calibrated, the student is referred to the !Am. Chem J.!, 16, 479.]
GENERAL DIRECTIONS FOR VOLUMETRIC ANALYSES
It cannot be too strongly emphasized that for the success of analyses uniformity of practice must prevail throughout all volumetric work with respect to those factors which can influence the accuracy of the measurement of liquids. For example, whatever conditions are imposed during the calibration of a burette, pipette, or flask (notably the time allowed for draining), must also prevail whenever the flask or burette is used.
The student should also be constantly watchful to insure parallel conditions during both standardization and analyst with respect to the final volume of liquid in which a titration takes place. The value of a standard solution is only accurate under the conditions which prevailed when it was standardized. It is plain that the standard solutions must be scrupulously protected from concentration or dilution, after their value has been established. Accordingly, great care must be taken to thoroughly rinse out all burettes, flasks, etc., with the solutions which they are to contain, in order to remove all traces of water or other liquid which could act as a diluent. It is best to wash out a burette at least three times with small portions of a solution, allowing each to run out through the tip before assuming that the burette is in a condition to be filled and used. It is, of course, possible to dry measuring instruments in a hot closet, but this is tedious and unnecessary.
To the same end, all solutions should be kept stoppered and away from direct sunlight or heat. The bottles should be shaken before use to collect any liquid which may have distilled from the solution and condensed on the sides.
The student is again reminded that variations in temperature of volumetric solutions must be carefully noted, and care should always be taken that no source of heat is sufficiently near the solutions to raise the temperature during use.
Much time may be saved by estimating the approximate volume of a standard solution which will be required for a titration (if the data are obtainable) before beginning the operation. It is then possible to run in rapidly approximately the required amount, after which it is only necessary to determine the end-point slowly and with accuracy. In such cases, however, the knowledge of the approximate amount to be required should never be allowed to influence the judgment regarding the actual end-point.
STANDARD SOLUTIONS
The strength or value of a solution for a specific reaction is determined by a procedure called !Standardization!, in which the solution is brought into reaction with a definite weight of a substance of known purity. For example, a definite weight of pure sodium carbonate may be dissolved in water, and the volume of a solution of hydrochloric acid necessary to exactly neutralize the carbonate accurately determined. From these data the strength or value of the acid is known. It is then a !standard solution!.
NORMAL SOLUTIONS
Standard solutions may be made of a purely empirical strength dictated solely by convenience of manipulation, or the concentration may be chosen with reference to a system which is applicable to all solutions, and based upon chemical equivalents. Such solutions are called !Normal Solutions! and contain such an amount of the reacting substance per liter as is equivalent in its chemical action to one gram of hydrogen, or eight grams of oxygen. Solutions containing one half, one tenth, or one one-hundredth of this quantity per liter are called, respectively, half-normal, tenth-normal, or hundredth-normal solutions.
Since normal solutions of various reagents are all referred to a common standard, they have an advantage not possessed by empirical solutions, namely, that they are exactly equivalent to each other. Thus, a liter of a normal solution of an acid will exactly neutralize a liter of a normal alkali solution, and a liter of a normal oxidizing solution will exactly react with a liter of a normal reducing solution, and so on.
Beside the advantage of uniformity, the use of normal solutions simplifies the calculations of the results of analyses. This is particularly true if, in connection with the normal solution, the weight of substance for analysis is chosen with reference to the atomic or molecular weight of the constituent to be determined. (See problem 26.)
The preparation of an !exactly! normal, half-normal, or tenth-normal solution requires considerable time and care. It is usually carried out only when a large number of analyses are to be made, or when the analyst has some other specific purpose in view. It is, however, a comparatively easy matter to prepare standard solutions which differ but slightly from the normal or half-normal solution, and these have the advantage of practical equality; that is, two approximately half-normal solutions are more convenient to work with than two which are widely different in strength. It is, however, true that some of the advantage which pertains to the use of normal solutions as regards simplicity of calculations is lost when using these approximate solutions.
The application of these general statements will be made clear in connection with the use of normal solutions in the various types of volumetric processes which follow.
I. NEUTRALIZATION METHODS
ALKALIMETRY AND ACIDIMETRY
GENERAL DISCUSSION
!Standard Acid Solutions! may be prepared from either hydrochloric, sulphuric, or oxalic acid. Hydrochloric acid has the advantage of forming soluble compounds with the alkaline earths, but its solutions cannot be boiled without danger of loss of strength; sulphuric acid solutions may be boiled without loss, but the acid forms insoluble sulphates with three of the alkaline earths; oxalic acid can be accurately weighed for the preparation of solutions, and its solutions may be boiled without loss, but it forms insoluble oxalates with three of the alkaline earths and cannot be used with certain of the indicators.
!Standard Alkali Solutions! may be prepared from sodium or potassium hydroxide, sodium carbonate, barium hydroxide, or ammonia. Of sodium and potassium hydroxide, it may be said that they can be used with all indicators, and their solutions may be boiled, but they absorb carbon dioxide readily and attack the glass of bottles, thereby losing strength; sodium carbonate may be weighed directly if its purity is assured, but the presence of carbonic acid from the carbonate is a disadvantage with many indicators; barium hydroxide solutions may be prepared which are entirely free from carbon dioxide, and such solutions immediately show by precipitation any contamination from absorption, but the hydroxide is not freely soluble in water; ammonia does not absorb carbon dioxide as readily as the caustic alkalies, but its solutions cannot be boiled nor can they be used with all indicators. The choice of a solution must depend upon the nature of the work in hand.
A !normal acid solution! should contain in one liter that quantity of the reagent which represents 1 gram of hydrogen replaceable by a base. For example, the normal solution of hydrochloric acid (HCl) should contain 36.46 grams of gaseous hydrogen chloride, since that amount furnishes the requisite 1 gram of replaceable hydrogen. On the other hand, the normal solution of sulphuric acid (H_{2}SO_{4}) should contain only 49.03 grams, i.e., one half of its molecular weight in grams.
A !normal alkali solution! should contain sufficient alkali in a liter to replace 1 gram of hydrogen in an acid. This quantity is represented by the molecular weight in grams (40.01) of sodium hydroxide (NaOH), while a sodium carbonate solution (Na_{2}CO_{3}) should contain but one half the molecular weight in grams (i.e., 53.0 grams) in a liter of normal solution.
Half-normal or tenth-normal solutions are employed in most analyses (except in the case of the less soluble barium hydroxide). Solutions of the latter strength yield more accurate results when small percentages of acid or alkali are to be determined.
INDICATORS
It has already been pointed out that the purpose of an indicator is to mark (usually by a change of color) the point at which just enough of the titrating solution has been added to complete the chemical change which it is intended to bring about. In the neutralization processes which are employed in the measurement of alkalies (!alkalimetry!) or acids (!acidimetry!) the end-point of the reaction should, in principle, be that of complete neutrality. Expressed in terms of ionic reactions, it should be the point at which the H^{+} ions from an acid[Note 1] unite with a corresponding number of OH^{-} ions from a base to form water molecules, as in the equation
H^{+}, Cl^{-} + Na^{+}, OH^{-} --> Na^{+}, Cl^{-} + (H_{2}O).
It is not usually possible to realize this condition of exact neutrality, but it is possible to approach it with sufficient exactness for analytical purposes, since substances are known which, in solution, undergo a sharp change of color as soon as even a minute excess of H^{+} or OH^{-} ions are present. Some, as will be seen, react sharply in the presence of H^{+} ions, and others with OH^{-} ions. These substances employed as indicators are usually organic compounds of complex structure and are closely allied to the dyestuffs in character.
[Note 1: A knowledge on the part of the student of the ionic theory as applied to aqueous solutions of electrolytes is assumed. A brief outline of the more important applications of the theory is given in the Appendix.]
BEHAVIOR OF ORGANIC INDICATORS
The indicators in most common use for acid and alkali titrations are methyl orange, litmus, and phenolphthalein.
In the following discussion of the principles underlying the behavior of the indicators as a class, methyl orange and phenolphthalein will be taken as types. It has just been pointed out that indicators are bodies of complicated structure. In the case of the two indicators named, the changes which they undergo have been carefully studied by Stieglitz (!J. Am. Chem. Soc.!, 25, 1112) and others, and it appears that the changes involved are of two sorts: First, a rearrangement of the atoms within the molecule, such as often occurs in organic compounds; and, second, ionic changes. The intermolecular changes cannot appropriately be discussed here, as they involve a somewhat detailed knowledge of the classification and general behavior of organic compounds; they will, therefore, be merely alluded to, and only the ionic changes followed.
Methyl orange is a representative of the group of indicators which, in aqueous solutions, behave as weak bases. The yellow color which it imparts to solutions is ascribed to the presence of the undissociated base. If an acid, such as HCl, is added to such a solution, the acid reacts with the indicator (neutralizes it) and a salt is formed, as indicated by the equation:
(M.o.)^{+}, OH^{-} + H^{+}, Cl^{-} --> (M.o.)^{+} Cl^{-} + (H_{2}O).
This salt ionizes into (M.o.)^{+} (using this abbreviation for the positive complex) and Cl^{-}; but simultaneously with this ionization there appears to be an internal rearrangement of the atoms which results in the production of a cation which may be designated as (M'.o'.)^{+}, and it is this which imparts a characteristic red color to the solution. As these changes occur in the presence of even a very small excess of acid (that is, of H^{+} ions), it serves as the desired index of their presence in the solution. If, now, an alkali, such as NaOH, is added to this reddened solution, the reverse series of changes takes place. As soon as the free acid present is neutralized, the slightest excess of sodium hydroxide, acting as a strong base, sets free the weak, little-dissociated base of the indicator, and at the moment of its formation it reverts, because of the rearrangement of the atoms, to the yellow form:
OH^{-} + (M'.o'.)^{+} --> [M'.o'.OH] --> [M.o.OH].
Phenolphthalein, on the other hand, is a very weak, little-dissociated acid, which is colorless in neutral aqueous solution or in the presence of free H^{+} ions. When an alkali is added to such a solution, even in slight excess, the anion of the salt which has formed from the acid of the indicator undergoes a rearrangement of the atoms, and a new ion, (Ph')^{+}, is formed, which imparts a pink color to the solution:
H^{+}, (Ph)^{-} + Na^{+}, OH^{-} --> (H_{2}O) + Na^{+}, (Ph)^{-} --> Na^{+}, (Ph')^{-}
The addition of the slightest excess of an acid to this solution, on the other hand, occasions first the reversion to the colorless ion and then the setting free of the undissociated acid of the indicator:
H^{+}, (Ph')^{-} --> H^{+}, (Ph)^{-} --> (HPh).
Of the common indicators methyl orange is the most sensitive toward alkalies and phenolphthalein toward acids; the others occupy intermediate positions. That methyl orange should be most sensitive toward alkalies is evident from the following considerations: Methyl orange is a weak base and, therefore, but little dissociated. It should, then, be formed in the undissociated condition as soon as even a slight excess of OH^{-} ions is present in the solution, and there should be a prompt change from red to yellow as outlined above. On the other hand, it should be an unsatisfactory indicator for use with weak acids (acetic acid, for example) because the salts which it forms with such acids are, like all salts of that type, hydrolyzed to a considerable extent. This hydrolytic change is illustrated by the equation:
(M.o.)^{+} C_{2}H_{3}O_{2}^{-} + H^{+}, OH^{-} --> [M.o.OH] + H^{+}, C_{2}H_{3}O_{2}^{-}.
Comparison of this equation with that on page 30 will make it plain that hydrolysis is just the reverse of neutralization and must, accordingly, interfere with it. Salts of methyl orange with weak acids are so far hydrolyzed that the end-point is uncertain, and methyl orange cannot be used in the titration of such acids, while with the very weak acids, such as carbonic acid or hydrogen sulphide (hydrosulphuric acid), the salts formed with methyl orange are, in effect, completely hydrolyzed (i.e., no neutralization occurs), and methyl orange is accordingly scarcely affected by these acids. This explains its usefulness, as referred to later, for the titration of strong acids, such as hydrochloric acid, even in the presence of carbonates or sulphides in solution.
Phenolphthalein, on the other hand, should be, as it is, the best of the common indicators for use with weak acids. For, since it is itself a weak acid, it is very little dissociated, and its nearly undissociated, colorless molecules are promptly formed as soon as there is any free acid (that is, free H^{+} ions) in the solution. This indicator cannot, however, be successfully used with weak bases, even ammonium hydroxide; for, since it is weak acid, the salts which it forms with weak alkalies are easily hydrolyzed, and as a consequence of this hydrolysis the change of color is not sharp. This indicator can, however, be successfully used with strong bases, because the salts which it forms with such bases are much less hydrolyzed and because the excess of OH^{-} ions from these bases also diminishes the hydrolytic action of water.
This indicator is affected by even so weak an acid as carbonic acid, which must be removed by boiling the solution before titration. It is the indicator most generally employed for the titration of organic acids.
In general, it may be stated that when a strong acid, such as hydrochloric, sulphuric or nitric acid, is titrated against a strong base, such as sodium hydroxide, potassium hydroxide, or barium hydroxide, any of these indicators may be used, since very little hydrolysis ensues. It has been noted above that the color change does not occur exactly at theoretical neutrality, from which it follows that no two indicators will show exactly the same end-point when acids and alkalis are brought together. It is plain, therefore, that the same indicator must be employed for both standardization and analysis, and that, if this is done, accurate results are obtainable.
The following table (Note 1) illustrates the variations in the volume of an alkali solution (tenth-normal sodium hydroxide) required to produce an alkaline end-point when run into 10 cc. of tenth-normal sulphuric acid, diluted with 50 cc. of water, using five drops of each of the different indicator solutions.
==================================================================== | | | | INDICATOR | N/10 | N/10 |COLOR IN ACID|COLOR IN ALKA- | H_{2}SO_{4}| NaOH |SOLUTION |LINE SOLUTION _______________|____________|__________|_____________|______________ | cc. | cc. | cc. | Methyl orange | 10 | 9.90 | Red | Yellow Lacmoid | 10 | 10.00 | Red | Blue Litmus | 10 | 10.00 | Red | Blue Rosalic acid | 10 | 10.07 | Yellow | Pink Phenolphthalein| 10 | 10.10 | Colorless | Pink ====================================================================
It should also be stated that there are occasionally secondary changes, other than those outlined above, which depend upon the temperature and concentration of the solutions in which the indicators are used. These changes may influence the sensitiveness of an indicator. It is important, therefore, to take pains to use approximately the same volume of solution when standardizing that is likely to be employed in analysis; and when it is necessary, as is often the case, to titrate the solution at boiling temperature, the standardization should take place under the same conditions. It is also obvious that since some acid or alkali is required to react with the indicator itself, the amount of indicator used should be uniform and not excessive. Usually a few drops of solution will suffice.
The foregoing statements with respect to the behavior of indicators present the subject in its simplest terms. Many substances other than those named may be employed, and they have been carefully studied to determine the exact concentration of H^{+} ions at which the color change of each occurs. It is thus possible to select an indicator for a particular purpose with considerable accuracy. As data of this nature do not belong in an introductory manual, reference is made to the following papers or books in which a more extended treatment of the subject may be found:
Washburn, E.W., Principles of Physical Chemistry (McGraw-Hill Book Co.), (Second Edition, 1921), pp. 380-387.
Prideaux, E.B.R., The Theory and Use of Indicators (Constable & Co., Ltd.), (1917).
Salm, E., A Study of Indicators, !Z. physik. Chem.!, 57 (1906), 471-501.
Stieglitz, J., Theories of Indicators, !J. Am. Chem. Soc.!, 25 (1903), 1112-1127.
Noyes, A.A., Quantitative Applications of the Theory of Indicators to Volumetric Analysis, !J. Am. Chem. Soc.!, 32 (1911), 815-861.
Bjerrum, N., General Discussion, !Z. Anal. Chem.!, 66 (1917), 13-28 and 81-95.
Ostwald, W., Colloid Chemistry of Indicators, !Z. Chem. Ind. Kolloide!, 10 (1912), 132-146.
[Note 1: Glaser, !Indikatoren der Acidimetrie und Alkalimetrie!. Wiesbaden, 1901.]
PREPARATION OF INDICATOR SOLUTIONS
A !methyl orange solution! for use as an indicator is commonly made by dissolving 0.05-0.1 gram of the compound (also known as Orange III) in a few cubic centimeters of alcohol and diluting with water to 100 cc. A good grade of material should be secured. It can be successfully used for the titration of hydrochloric, nitric, sulphuric, phosphoric, and sulphurous acids, and is particularly useful in the determination of bases, such as sodium, potassium, barium, calcium, and ammonium hydroxides, and even many of the weak organic bases. It can also be used for the determination, by titration with a standard solution of a strong acid, of the salts of very weak acids, such as carbonates, sulphides, arsenites, borates, and silicates, because the weak acids which are liberated do not affect the indicator, and the reddening of the solution does not take place until an excess of the strong acid is added. It should be used in cold, not too dilute, solutions. Its sensitiveness is lessened in the presence of considerable quantities of the salts of the alkalies.
A !phenolphthalein solution! is prepared by dissolving 1 gram of the pure compound in 100 cc. of 95 per cent alcohol. This indicator is particularly valuable in the determination of weak acids, especially organic acids. It cannot be used with weak bases, even ammonia. It is affected by carbonic acid, which must, therefore, be removed by boiling when other acids are to be measured. It can be used in hot solutions. Some care is necessary to keep the volume of the solutions to be titrated approximately uniform in standardization and in analysis, and this volume should not in general exceed 125-150 cc. for the best results, since the compounds formed by the indicator undergo changes in very dilute solution which lessen its sensitiveness.
The preparation of a !solution of litmus! which is suitable for use as an indicator involves the separation from the commercial litmus of azolithmine, the true coloring principle. Soluble litmus tablets are often obtainable, but the litmus as commonly supplied to the market is mixed with calcium carbonate or sulphate and compressed into lumps. To prepare a solution, these are powdered and treated two or three times with alcohol, which dissolves out certain constituents which cause a troublesome intermediate color if not removed. The alcohol is decanted and drained off, after which the litmus is extracted with hot water until exhausted. The solution is allowed to settle for some time, the clear liquid siphoned off, concentrated to one-third its volume and acetic acid added in slight excess. It is then concentrated to a sirup, and a large excess of 95 per cent. alcohol added to it. This precipitates the blue coloring matter, which is filtered off, washed with alcohol, and finally dissolved in a small volume of water and diluted until about three drops of the solution added to 50 cc. of water just produce a distinct color. This solution must be kept in an unstoppered bottle. It should be protected from dust by a loose plug of absorbent cotton. If kept in a closed bottle it soon undergoes a reduction and loses its color, which, however, is often restored by exposure to the air.
Litmus can be employed successfully with the strong acids and bases, and also with ammonium hydroxide, although the salts of the latter influence the indicator unfavorably if present in considerable concentration. It may be employed with some of the stronger organic acids, but the use of phenolphthalein is to be preferred.
PREPARATION OF STANDARD SOLUTIONS
!Hydrochloric Acid and Sodium Hydroxide. Approximate Strength!, 0.5 N
Comments
Log in to leave a comment.